Product Description
Ethylene oxide (also known as EO or EtO) is an alkylating agent that disrupts cellular metabolism and reproductive processes of microorganisms.
Basing on vacuum processes, EO penetrates breathable packaging, making contact with all accessible surfaces of the product, kill microorganisms. and achieve the required sterility assurance level (SAL).
Ethylene oxide sterilization consists of four primary variables: gas concentration, humidity, temperature, and time.The low temperature process makes EO an ideal method for a wide variety of healthcare products, including masks, protective clothing, and electronic medical instruments, etc.
Parameters
Product form | Manual door cabinet | Working temperature | 45℃-55℃ |
Heating method | Electric heating (AC380V, 50Hz) | Working pressure | -50kpa |
Material | 304 anti-corrosion stainless steel | Highest anti-stress value | -80kpa |
Sealing door material | 20mm thick 5052 aluminum | Humidity | 45% -75% |
Selected gas | 80% EO, 20% CO2 mixed gas (0.8-1.2g/L) | ||
Sterilization time | 0-99 hours adjustable | ||
Vacuum speed | pumped to -60kpa in 25 minutes | ||
Air replacement | Automatic ventilation, the number and interval of ventilation can be set freely | ||
Display status | Display temperature, humidity, pressure, sterilization time, residual gas treatment time, times | ||
Sensor | Humidity Sensor;Pressure Sensor;Temperature Sensor | ||
Humidification System | Pulse humidification under vacuum; | ||
Residual gas treatment device | The filtration rate of the filter to the particle diameter ≥0.3μm is not less than 99.5%. | ||
Leakage rate of sterilization chamber | The measurement of the leakage rate is carried out under the conditions of no load, the temperature of the sterilization chamber is unchanged, the test pressure is -50kPa, and the measurement time is 1h. The average leakage rate within the measurement time is ≤0.1kPa / min. | ||
Vacuum rate | Under no-load conditions, the time for the sterilization cabinet to reach -60kPa from atmospheric pressure should be ≤30min. | ||
Multiple protection functions | over-voltage and over-temperature protection functions high temperature alarm system and ultra-low pressure alarm system. |
Applications
EO is suitable for the sterilization of a wide range of materials not compatible with other methods of sterilization. Thereforce, it is widely used in mask factories, hospital CSSD, dental clinic ,ophthalmology clinic, pet hospitals, and test labs.